Device evaluated by mfr.: the product return contained an angiojet solent omni.The device was returned inside of the packaging, with the white strip that indicates a sealed back missing.The assembly, supply line, shaft, saline supply and tip were all visually examined for potential damage.A functional test was performed.The pump was placed into the ultra drive console and the device primed with no complication.The device was run for a total time of 90 seconds in thrombectomy more.The device stayed within the expected range of the product, with an average pressure.No damage or discrepancies were noted on the rest of the catheter upon visual inspection.
|