• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPTIFLEX; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPTIFLEX; SNARE, FLEXIBLE Back to Search Results
Model Number M00562402
Device Problems Failure to Deliver Energy (1211); Entrapment of Device (1212); Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/06/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
Note: this report pertains to one of two devices used during the same procedure.Refer to manufacturer report # 3005099803-2021-04530 for the associated device information.It was reported to boston scientific corporation that a captiflex medium oval flexible snare was used during a colonoscopy procedure performed on (b)(6) 2021.During the procedure, an oval snare was unable to cauterize a very large rectal polyp.The snare became embedded in the polyp and was cut to remove.This occurred with two snares from the same lot number.Two different cautery machines were tried and the snares did not work with either.The snares were tested before placing them around the polyp and worked at that time.Reportedly, the snares were securely attached to the active cord and there were no visible issues noted with the cautery pins.The procedure was completed with a different device and taking the polyp in small pieces.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIFLEX
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key12414894
MDR Text Key271671129
Report Number3005099803-2021-04531
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729501640
UDI-Public08714729501640
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131700
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/06/2024
Device Model NumberM00562402
Device Catalogue Number6240-40
Device Lot Number0027085177
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/11/2021
Initial Date FDA Received09/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-