Model Number M00562402 |
Device Problems
Failure to Deliver Energy (1211); Entrapment of Device (1212); Failure to Cut (2587)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/06/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
Note: this report pertains to one of two devices used during the same procedure.Refer to manufacturer report # 3005099803-2021-04530 for the associated device information.It was reported to boston scientific corporation that a captiflex medium oval flexible snare was used during a colonoscopy procedure performed on (b)(6) 2021.During the procedure, an oval snare was unable to cauterize a very large rectal polyp.The snare became embedded in the polyp and was cut to remove.This occurred with two snares from the same lot number.Two different cautery machines were tried and the snares did not work with either.The snares were tested before placing them around the polyp and worked at that time.Reportedly, the snares were securely attached to the active cord and there were no visible issues noted with the cautery pins.The procedure was completed with a different device and taking the polyp in small pieces.There were no patient complications reported as a result of this event.
|
|