ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problems Unstable (1667); Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/24/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Device failed the rpm test and it was noticed there was a leak around the swivel.The control bar was flush with the master blade and therefore does meet standards.Calibration was not in limits when set to the 0 reading.The device did not pass visual inspection.It needed an aged repair which the customer declined.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
It was reported the device was sent for preventative maintenance and found in need of repair.Investigation found that the device failed the rpm test and it was noticed there was a leak around the swivel.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key12416722
MDR Text Key269729808
Report Number0001526350-2021-00986
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number2848300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2021
Was the Report Sent to FDA? No
Date Manufacturer Received09/01/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-