MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C1154KP
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2021
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a small volume multirate infusor was torn.This was observed prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Device manufactured between october 22, 2019 to october 23, 2019.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
The device was received for evaluation.Visual inspection was performed using the naked eye which observed a large hole (torn) in the primary packaging.The device inside the packaging was examined and no issues were noted.The reported condition was verified.The cause of the condition could not be determined; however, the most probable cause may potentially be related to the handling of the product during unpacking.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key12418270
MDR Text Key270181806
Report Number1416980-2021-05424
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412081397
UDI-Public(01)00085412081397
Combination Product (y/n)Y
PMA/PMN Number
K011317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C1154KP
Device Lot Number19K028
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2021
Date Manufacturer Received10/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-