Catalog Number 2C1154KP |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a small volume multirate infusor was torn.This was observed prior to patient use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Device manufactured between october 22, 2019 to october 23, 2019.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
The device was received for evaluation.Visual inspection was performed using the naked eye which observed a large hole (torn) in the primary packaging.The device inside the packaging was examined and no issues were noted.The reported condition was verified.The cause of the condition could not be determined; however, the most probable cause may potentially be related to the handling of the product during unpacking.Should additional relevant information become available, a supplemental report will be submitted.
|
|