Model Number 295025-001 |
Device Problems
Battery Problem (2885); No Apparent Adverse Event (3189)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/23/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The freedom onboard battery will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The customer, a syncardia authorized distributor, reported that patient's freedom onboard battery discharged immediately as soon as it was inserted into the freedom driver after a complete charge.
|
|
Manufacturer Narrative
|
The freedom onboard battery was evaluated per the freedom onboard battery evaluation procedure and passed all sections.During investigation testing, the customer-reported issue was not reproduced and there was no evidence of a device malfunction.The freedom driver as well as the freedom battery charger used at the time of the reported event were not returned and therefore could not be included in this investigation.The root cause of the customer-reported issue that the freedom onboard battery discharged immediately after it was inserted into the freedom driver after a complete charge could not be conclusively determined.The battery performed as intended.This issue will be tracked and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(6).(b)(4)follow-up report 1.
|
|