SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problems Computer Software Problem (1112); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2021
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
 
Event Description
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the companion 2 driver did not boot up with the proper user interface screen.
 
Manufacturer Narrative
The patient data file was reviewed and no alarms were recorded that would indicate a device malfunction.During functional testing, the key was observed to be hard to turn and had to be forced to the "on" position.With the key switch faulty and not allowing proper boot up, it is likely that the left and right embedded supervisors were not properly communicating which could lead to improper boot up issues to include a white screen as reported by the customer.The root cause of the customer-reported issue was determined to be a malfunction of the companion 2 driver key switch.Syncardia has a corrective and preventive action (capa) for the issue of companion 2 driver broken keys/key switch.Syncardia has completed its investigation and is closing this file.(b)(4) follow- up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
sagar pimpalwar
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key12433822
MDR Text Key273806855
Report Number3003761017-2021-00179
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/25/2021
Initial Date FDA Received09/07/2021
Supplement Dates Manufacturer Received08/25/2021
Supplement Dates FDA Received09/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/06/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-