Catalog Number 110003450 |
Device Problem
Fracture (1260)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
A g7 shell inserter was returned and the instrument tip was determined to be fracture.No adverse consequences were reported as a result of this malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.One g7 str monoblock shell insrtr item# 110003450 lot# 049300 was returned and evaluated.Upon visual inspection there was some scuffing to the strike plate of the device.The threaded tip of the device had fractured.The common failure mode for the threaded inserter tips presented in this summary is bending overload of the threaded tip, some of which may have also had minimal threads engaged at the time of fracture.Device history record (dhr) was reviewed and no discrepancies were found.Root cause of the event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|