Catalog Number 110003450
Device Problem Fracture (1260)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
A g7 shell inserter was returned and the instrument tip was determined to be fracture.No adverse consequences were reported as a result of this malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.One g7 str monoblock shell insrtr item# 110003450 lot# 049300 was returned and evaluated.Upon visual inspection there was some scuffing to the strike plate of the device.The threaded tip of the device had fractured.The common failure mode for the threaded inserter tips presented in this summary is bending overload of the threaded tip, some of which may have also had minimal threads engaged at the time of fracture.Device history record (dhr) was reviewed and no discrepancies were found.Root cause of the event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12435976
MDR Text Key270306164
Report Number0001825034-2021-02618
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number110003450
Device Lot Number049300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/11/2021
Initial Date FDA Received09/07/2021
Supplement Dates Manufacturer Received12/09/2021
Supplement Dates FDA Received12/10/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-