The patient was undergoing a coil embolization procedure in the ovarian vein using pod packing coils (pod pcs).During the procedure, the hospital resident had kinked the pusher assembly of a pod pc while advancing the coil within the introducer sheath, and the pod pc detached.Therefore, the pod pc was removed.The procedure was completed using a new pod pc.There was no report of an adverse effect to the patient.
|
Please note that the following section was inadvertently missed on the initial mfr report and is being updated on this follow-up #01 mfr report:3005168196-2021-01970.1.Section h.Box 6.Device code 2.Evaluation of the returned pod pc confirmed that the pusher assembly was kinked in multiple locations throughout its length, and the embolization coil was detached.Evaluation revealed that the pet lock was intact on the proximal end of the pusher assembly, and pull wire was within ddt.If the pod pc forcefully advanced against resistance, damage such as a kink may occur.Subsequently, if the device is retracted against resistance, the pusher assembly may elongate beyond the reach of the pull wire and result in the embolization coil detaching from the pusher assembly.Penumbra products are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|