MAUDE Adverse Event Report: BIOMET UK LTD. OXF CMNTLS IMPLANT INSERT; OXFORD MICROPLASTY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF CMNTLS IMPLANT INSERT; OXFORD MICROPLASTY INSTRUMENT Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Initial report.Report source, foreign - event occurred in the (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported, that one of the lips of the instrument that fixate the tibia component on the inserter broke off.Attempts have been made but no further information has been provided at this time.
 
Manufacturer Narrative
(b)(4), this follow-up report is being submitted to relay further information.The event occurred (b)(6), 2021.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that one of the lips of the instrument that fixate the tibia component on the inserter was broken off.
 
Event Description
It was reported that one of the lips of the instrument that fixate the tibia component on the inserter was broken off.
 
Manufacturer Narrative
(b)(4).This final report is being submitted to relay additional information.Complaint summary: as the product has not been received, the investigation was limited to the information provided; a review of device history records and complaint history.In addition, we have not been provided with any supporting documentation which could provide additional information.A review of the manufacturing history records confirms no abnormalities or deviations reported.A review of the complaint database over the last 3 years has found 34 similar complaints identified with this item number including (b)(4).Without the opportunity to examine the complaint product, the root cause cannot be determined due to insufficient information.Capa: no corrective or preventive action required at this time.If any further information is found which would change or alter any conclusions or information, a supplemental will be submitted accordingly.Zimmer biomet will continue to monitor for trends.H3 other text : item discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF CMNTLS IMPLANT INSERT
Type of Device
OXFORD MICROPLASTY INSTRUMENT
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12437478
MDR Text Key271650220
Report Number3002806535-2021-00382
Device Sequence Number1
Product Code NRA
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number32-422097
Device Lot NumberZB151201
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/31/2021
Initial Date FDA Received09/08/2021
Supplement Dates Manufacturer Received09/24/2021
12/13/2021
Supplement Dates FDA Received10/04/2021
12/14/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-