SIEMENS MEDICAL SOLUTIONS SOMATOM SENSATION 64; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 8377520
Device Problems Device Displays Incorrect Message (2591); Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 06/23/2021
Event Type  malfunction  
Event Description
During the patient's ct scan, the ct machine briefly paused and caused a delay in the timing of the scan.Siemens was contacted and a service request was placed.Corrective action/service performed : replaced vega p/s.System still aborts checkup and cal with dom_3003 and xdc-3003 errors.Will continue work on 2nd notification for same fl.Task remains open for possible rsc evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM SENSATION 64
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
SIEMENS MEDICAL SOLUTIONS
40 liberty boulevard
mailcode: 64-1a
malvern PA 19355
MDR Report Key12438230
MDR Text Key270287983
Report Number12438230
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 08/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8377520
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA08/31/2021
Event Location Hospital
Date Report to Manufacturer09/08/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age25915 DA
-
-