Model Number 71940-01 |
Device Problems
Unable to Obtain Readings (1516); Material Twisted/Bent (2981)
|
Patient Problems
Hypoglycemia (1912); Loss of consciousness (2418)
|
Event Date 07/19/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered in is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A bent sensor tip was reported with the freestyle libre sensor and customer was therefore unable to apply the device and monitor glucose levels.As a result, the customer experienced a loss of consciousness and was unable to self-treat, requiring treatment of sugar by a third-party.There was no report of death or permanent injury associated with this event.
|
|
Event Description
|
A bent sensor tip was reported with the freestyle libre sensor and customer was therefore unable to apply the device and monitor glucose levels.As a result, the customer experienced a loss of consciousness and was unable to self-treat, requiring treatment of sugar by a third-party.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Additional information: section d4 (expiration date) and h4 (device mfg date) were updated based on the extended investigation.No product has been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specifications.The dhrs (device history review) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.If the product is returned, the case will be re-opened, and a physical investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
|
|