Catalog Number 8606000 |
Device Problems
Gas Output Problem (1266); Failure to Deliver (2338); Pressure Problem (3012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
For the investigation the provided logfile was analyzed.It was found that the auxiliary vacuum pressure dropped below the minimum value for safe operation.The auxiliary vacuum pressure is required to actuate the valves that control the ventilation and to keep the diaphragm of the ventilator piston in place to avoid wrinkling during piston movement.Insufficient vacuum pressure may result in serious damages to the ventilator unit and thus, the device is designed to shut down automatic ventilation upon occurrence.As reported, the ventilator shut-down was alerted to the user by means of a corresponding alarm.The user can switch to manual ventilation as described in the instructions for use.Monitoring is still functional.Potential root causes for the drop in vacuum pressure are manifold and are not necessarily related to device malfunctions - also an accidental disconnected actuator tube of the apl bypass valve can cause these symptoms.
|
|
Event Description
|
It was reported that an acoustical and visual "ventilator failure" alarm was posted during use on patient.No injury reported.
|
|
Event Description
|
It was reported that an acoustical and visual "ventilator failure" alarm was posted during use on patient.No injury reported.
|
|
Manufacturer Narrative
|
Due to a technical issue with our internal emdr system we submitted for the form fda 3500a an incorrect value for the field h3.- not returned to manufacturer.
|
|
Search Alerts/Recalls
|