Model Number 306546
Device Problem Failure to Deliver (2338)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: it was reported the plunger would not flush past a certain point.To aid in the investigation, three samples, each in a plastic bag, with no packaging flow wrap and one photo was received for evaluation by our quality team.A visual inspection was performed and no defects or imperfections were observed.Each sample was then tested for sustaining force, which is the force applied to the plunger rod while moving downwards when expelling the saline solution, and all results were within specification.The photo provided shows the samples received.It could be possible that the customer is getting some products that are towards the high specification limit and are related to the symptom reported by the customer since they require extra force than normal to expel the solution.A device history record review was completed for provided material number 306546, lot numbers 1019137 and 1035833.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.Previous investigations have revealed that improper silicone application within the syringe barrel can create plunger resistance.Several quality initiatives have been implemented on our manufacturing line to ensure that the silicone application is properly applied during the manufacturing process.Additionally, a monitoring program is also in place to verify the silicone is uniformly applied to the syringe barrel.Based on the investigation and with the returned sample analysis the symptom reported by the customer of plunger movement difficult is confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that syringe 10ml reg pr saline 10ml fill had a difficult to move plunger.The following information was provided by the initial reporter: "the plunger would only flush up to approx the 3 or 4ml line and would not advance.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 10ML REG PR SALINE 10ML FILL
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12449062
MDR Text Key270634329
Report Number1911916-2021-00967
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public30382903065463
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/31/2024
Device Model Number306546
Device Catalogue Number306546
Device Lot Number1035833
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/11/2021
Initial Date FDA Received09/09/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-