• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER ACCESS SARS-COV-2 IGG II; IN VITRO DIAGNOSTIC DEVICES: CORONAVIRUS SEROLOGICAL REAGENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER ACCESS SARS-COV-2 IGG II; IN VITRO DIAGNOSTIC DEVICES: CORONAVIRUS SEROLOGICAL REAGENTS Back to Search Results
Catalog Number C69057
Device Problem False Negative Result (1225)
Patient Problem Insufficient Information (4580)
Event Date 08/26/2021
Event Type  malfunction  
Event Description
On (b)(6) 2021 the customer reported a non-reactive covid igg result (access sars-cov-2 igg ii, part number c69057, lot number not provided) was generated on the customer's dxi (unicel dxi 800 access immunoassay analyzer, part number 973100 and serial number (b)(4)) on (b)(6) 2021 for a patient who had been previously vaccinated against covid.The customer reported the patient had also tested positive for covid using a pcr assay (assay information not provided) the previous year.The customer did not indicate whether the results were released from the laboratory.The customer did not report a change to patient care of treatment in connection with this event.The customer reported that the patient had been vaccinated against sars-cov-2; however, customer did not provide vaccination dates or specific vaccination administered.Customer reported the patient had a reactive covid result using an alternate platform, coviprotect sars cov2 spike antibody quantitative.No hardware errors were reported in conjunction with this event.System performance indicators such as system check, calibration and quality control not provided for review.No issues with sample integrity were reported by the customer.Sample collection, handling and processing information such as sample type, sample volume collected, sample quality, centrifugation time and speed, storage temperature and other information was not provided by the customer.
 
Manufacturer Narrative
(b)(6).The customer did not provide patient demographics such as date of birth, weight, ethnicity or race.Customer did not provide a reagent lot number; therefore, an expiration date and udi could not be provided.Initial reporter name and telephone number were not provided.Customer did not provide a reagent lot number; therefore, a date of manufacture could not be provided.The access sars-cov-2 igg ii assay was not returned for evaluation.There were no reports of system issues at the time of the event.There was no report of issues with other assays at the time of the event.No hardware errors or flags were reported in conjunction with the event.Sars-cov-2 is an enveloped non-segmented positive-sense rna virus.It has several structural proteins including spike (s), envelope (e), membrane (m) and nucleocapsid (n).The spike protein (s) contains a receptor binding domain (rbd) which is responsible for recognizing the cell surface receptor, angiotensin converting enzyme-2 (ace2).It is found that the rbd of the sars-cov-2 s protein strongly interacts with the human ace2 receptor leading to endocytosis into the host cells and viral replication.The access assay detects antibodies directed against the spike protein, which are more likely to neutralize the virus.No manufacturer guarantees both a specificity and sensitivity of 100%.The concentration of sars-cov-2 igg in a given specimen determined with assays from different manufacturers can vary due to differences in assay methods, diversity of antibodies and reagent specificity.Differences in each individual assay are expected.Different vaccines do not elicit the same immune response and each patient response is different therefore differences in patient responses are expected after vaccination.The access assay is not labeled for vaccine response detection.Additionally, per safety communication from fda on 19may2021, antibody testing is not currently recommended to assess immunity after covid-19 vaccination.In conclusion, the cause of this event cannot be determined with the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS SARS-COV-2 IGG II
Type of Device
IN VITRO DIAGNOSTIC DEVICES: CORONAVIRUS SEROLOGICAL REAGENTS
Manufacturer (Section D)
BECKMAN COULTER
1000 lake hazeltine drive
chaska MN 55318
MDR Report Key12450902
MDR Text Key281159922
Report Number2122870-2021-00138
Device Sequence Number1
Product Code QKO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 09/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC69057
Was Device Available for Evaluation? No
Date Manufacturer Received08/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-