Catalog Number 2C8541
Device Problems Filling Problem (1233); Material Twisted/Bent (2981); No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a clearlink system continu-flo solution set under infused.This occurred during infusion with calcium chloride.It was stated that the new bag hung only had about 100 ml gone where almost 500 ml should have infused.When the tubing was taken out of the pump it was noted it was completely flat and occluded.It was further reported that the unspecified pump never alarmed and stated the volume had infused.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional test including pressure test and clear passage test was performed with no defects observed.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK DUO-VENT CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
MDR Report Key12454151
MDR Text Key270840686
Report Number1416980-2021-05590
Device Sequence Number1
Product Code FPA
Combination Product (y/n)Y
PMA/PMN Number
K961225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 10/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C8541
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2021
Date Manufacturer Received10/12/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CALCIUM CHLORIDE; UNSPECIFIED PUMP; CALCIUM CHLORIDE; UNSPECIFIED PUMP
Patient Outcome(s) Congenital Anomaly; Disability;
-
-