• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2021
Event Type  malfunction  
Event Description
It was reported that patient presented in clinic for implant procedure.During procedure, it was difficult to remove the guidewire from the left ventricular (lv) lead and the header pin of the lv lead was found to be broken and detached.The lv lead was removed and replaced on (b)(6) 2021.The patient was discharged.
 
Manufacturer Narrative
The reported events of header pin broken and difficulty removing the guidewire were confirmed by analysis.Final analysis found deformation of inner coil and separation connector cap consistent with procedural damage.No other anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key12457666
MDR Text Key270990214
Report Number2017865-2021-29916
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000110375
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2021
Was the Report Sent to FDA? No
Date Manufacturer Received09/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age85 YR
Patient Weight99
-
-