MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problems Gas/Air Leak (2946); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Air Embolism (1697)
Event Date 08/18/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, after exchanging from a competitor sheath to the sheath for the cryo procedure, air was observed in the left atrium and aorta near the brachiocephalic artery.When the dilator was removed and aspiration was attempted, air was pulled into the syringe.The case was aborted.A brain scan was performed and no anomalies were observed.The patient was reported to be stable post-procedure.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the sheath 4fc12 with lot number 0010680960 was returned and analyzed.Visual inspection showed that the sheath was intact and had no apparent issues.The deflection mechanism functioned as per specification.The reported air ingress issue was not reproduced during testing.A leak test for the sheath was done on the sentinel blackbelt tester, the pressure decay was within an acceptable limit.The shaft and the valve were leak tight with no apparent issues.Several aspiration/flushing tests were performed, and they did not show any air passing through the tube or expelled from the sheath distal tip.In conclusion, the clinical issues (air embolism and st elevation) could not be confirmed through testing.The reported air ingress could not be confirmed through testing.The balloon catheter passed the returned product inspection as per specifications.There is no indication of relation of adverse events to the performance of the product.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key12460758
MDR Text Key271055723
Report Number3002648230-2021-00428
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/19/2023
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0010680960
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2021
Date Manufacturer Received09/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-