MAUDE Adverse Event Report: ABBOTT IRELAND ARCHITECT ANTI-HBS REAGENT KIT; TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND ARCHITECT ANTI-HBS REAGENT KIT; TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM) Back to Search Results
Catalog Number 07C18-75
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/28/2021
Event Type  malfunction  
Manufacturer Narrative
This report is being filed on an international product, list number 07c18 that has a similar product distributed in the us, list number 01l82.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer observed (b)(6) architect (b)(6) results for one male patient.The initial result (b)(6) result was (b)(6) during a physical examination test.Two days later, a new blood test was taken, and the results were (b)(6).No impact to patient management was reported.
 
Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review & device history record review.Return testing was not completed as returns were not available.A review of tracking and trending did not identify any trends for the complaint issue.Device history record review did not identify any issues associated with lot 19676fn01 and the complaint issue.A review of the labelling and literature addresses the customer¿s issue.Per product labelling, if the anti-hbs results are inconsistent with clinical evidence, additional testing is suggested to confirm the result.For diagnostic purposes, results should be used in conjunction with patient history and other hepatitis markers for diagnosis of acute, chronic, or recovered infection.In this case, negative results or lower results were obtained on retest on another instrument, indicating possible sample or reagent handling or integrity issues or instrumentation issues for the original testing.The overall specificity for the architect anti-hbs assay (determined by considering result values of = 10.00 miu/ml as reactive) was estimated to be 99.22% at the lower 95% confidence level, therefore false reactive results may at times occur.False positive results may arise as a result of sample or reagent integrity issues at time of testing and details regarding sample and reagent handling are provided within the product package insert.Based on the investigation, no systemic issue or deficiency of the architect anti-hbs reagent lot 19676fn01 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT ANTI-HBS REAGENT KIT
Type of Device
TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM)
Manufacturer (Section D)
ABBOTT IRELAND
finisklin business park
sligo F91VY 44
EI  F91VY44
MDR Report Key12462773
MDR Text Key271122131
Report Number3008344661-2021-00167
Device Sequence Number1
Product Code LOM
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 10/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/26/2021
Device Catalogue Number07C18-75
Device Lot Number19676FN01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/31/2021
Initial Date FDA Received09/13/2021
Supplement Dates Manufacturer Received10/05/2021
Supplement Dates FDA Received10/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARC I1000SR MOD, 01L86-01, I1SR56695; ARC I1000SR MOD, 01L86-01, I1SR56695; ARC I1000SR MOD, (B)(4)
-
-