The product lot number was not provided, therefore, the manufacturing records could not be reviewed.This report is associated with mfr report number: 3005168196-2021-02026, 3005168196-2021-02027, 3005168196-2021-02028.
|
The patient was undergoing a thrombectomy procedure in the middle cerebral artery (mca) using a penumbra engine (engine), a penumbra engine canister (canister), aspiration tubing (tubing), and a penumbra system ace 68 reperfusion catheter (ace68).During the procedure, some of the engine indicator lights did not illuminate when the device was powered on.The technologist exchanged the canister for a new one and attempted to re-seat it in the engine; however, the issue persisted.Therefore, the canister was removed and the tubing was exchanged for a new one.While using the new tubing and a new canister, only two of the engine indicator lights illuminated.Therefore, the engine and the canister were removed from the procedure.The procedure was completed using a new engine, a new canister, the same tubing, and the same ace68.There was no report of an adverse effect to the patient.
|