MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problem Firing Problem (4011)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #: v94c39.Date of event is 2021.Event day and event month were not reported.Investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Upon visual analysis of the returned sample, it was determined the el5ml device was returned with no damage in the external components.Upon visual inspection, one clip was noted to be partially formed in the jaws.The condition of the clips is most likely due to the trigger not completely "squeezing." the clip was removed in order to perform functional testing.In an attempt to replicate the reported incident, the instrument was tested for functionality.Upon testing, the device was cycled, fed, and formed the remaining nine clips as intended, then the device locked out as intended, but the orange indicator did not show up.This finding is not related to the incident reported.In order to evaluate the condition of the internal components the device was disassembled.Upon disassembly, no anomalies were found.Please note that this condition is unrelated to the reported incident.The partially formed clip in the jaws may have caused the device to not fire the clips properly or at all, however, no conclusion could be reached as to what may have caused the clip and the condition of the indicator.Device was fired without any difficulty.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following: "caution: insert the clip applier through an appropriately-sized trocar.The empty jaws will passively collapse as they are inserted through a 5mm trocar and reopen when completely through the trocar.Prior to positioning the jaws around the tubular structure or vessel, load a clip into the jaws by partially squeezing the trigger in a smooth continuous motion for approximately one-third of the total firing stroke.Prior to loading a clip in the jaws and firing the instrument: ensure that the jaws are fully open by verifying that the line of demarcation between the jaws and the instrument shaft is past the distal end of the trocar cannula.Prior to positioning the jaws around the tubular structure or vessel, load a clip into the jaws by partially squeezing the trigger in a smooth continuous motion for approximately one-third of the total firing stroke.Position the jaws with the preloaded clip completely around the tubular structure or vessel to be ligated.The structure to be ligated should be positioned against the apex of the clip." as part of our quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event Description
It was reported that during an unknown procedure, the clips were not applying properly.It is unknown how the procedure was completed.There was no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo
*  
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
*  
6107428552
MDR Report Key12465715
MDR Text Key271253063
Report Number3005075853-2021-05404
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 08/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberV94C39
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2021
Initial Date Manufacturer Received 08/26/2021
Initial Date FDA Received09/13/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-