Model Number 71440311
Device Problems Break (1069); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2021
Event Type  malfunction  
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned gii mod pat reamer gde confirms the pin that holds the device together in the middle is broken.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Event Description
It was reported that the supporting arm of a genesis ii modified patellar reamer guide is broken.In addition, it was noted that the pin fell out.As this was noticed in a non-surgical environment, there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII MOD PAT REAMER GDE
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12472081
MDR Text Key271459766
Report Number1020279-2021-06859
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010462749
UDI-Public03596010462749
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440311
Device Catalogue Number71440311
Device Lot Number16BBG0042
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/23/2021
Initial Date FDA Received09/14/2021
Supplement Dates Manufacturer Received09/14/2021
Supplement Dates FDA Received09/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/13/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-