SYNTHES GMBH 3.2MM DRILL BIT/QC/145MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, PNEUMATIC POWERED & ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 310.310
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2021
Event Type  malfunction  
Manufacturer Narrative
Additional patient identifier: (b)(6).Additional device product codes: hwe; gfa; gff.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that during a procedure on (b)(6) 2021, the drill bit broke.Nothing is left in the patient's bone.The procedure was successfully completed with a backup device.There was no surgical delay and no other medical intervention required.This report is for (1) 3.2mm drill bit/qc/145mm this is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6: customer quality investigation: the device was not returned.A photo-investigation was performed on the image provided.Upon inspecting the image provided, the complaint condition was confirmed as the proximal end of the drill bit was observed to be broken.However, a definitive assignable root cause of cannot be determined based on the available image.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.Conclusion: the complaint condition can be confirmed during photo investigation.During the investigation, no product design issues, or discrepancies were observed.No manufacturing issues were noted during investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.H6: a device history record (dhr) review was conducted: part: 310.310, lot: 119p779, manufacturing site: bettlach, release to warehouse date: 29 april 2021.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The drill bit ø3.2 l145/120 2flute (part# 310.310, lot# 4119p779 was returned and received at us cq.Upon visual inspection at cq, it is observed that the proximal end of the drill bit was broken and broken piece was returned.No other issues were identified with the returned device.The complaint is confirmed for the received device.There is no indication that a design or manufacturing issue contributed to the complaint.While no definitive root cause could be determined it is possible that the device encountered unintended forces during its use.No new malfunctions were observed during this investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history lot : part: 310.310 lot: 119p779 manufacturing site: bettlach release to warehouse date: 29 april 2021 a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.2MM DRILL BIT/QC/145MM
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, PNEUMATIC POWERED & ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key12477566
MDR Text Key271579738
Report Number8030965-2021-07767
Device Sequence Number1
Product Code HSZ
UDI-Device Identifier07611819018693
UDI-Public(01)07611819018693
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number310.310
Device Lot Number119P779
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/18/2021
Initial Date FDA Received09/15/2021
Supplement Dates Manufacturer Received09/28/2021
11/10/2021
Supplement Dates FDA Received09/29/2021
11/11/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age13 YR