Model Number MC2-5570R |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
Failure of Implant (1924)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
|
Event Description
|
Information was received that the rod could not be distracted anymore.No patient adverse event was reported.
|
|
Manufacturer Narrative
|
The returned rod was observed to be partially distracted with some score marks on the distraction rod indicating the rod was partially distracted.The rod was functionally tested and was able to distract and retract with the manual distractor.The rod was able to retract/distract with fast distractor to perform stroke test the maximum and minimum length were measured at 303.43mm and 275.87mm respectively; giving a total stroke measurement of 27.56mm, which meets the 27-29mm specification.The work order was reviewed and confirmed the rod passed all inspections.Review of the device history records revealed no discrepancies related to this complaint.The rod was manufactured in accordance with the specified requirements and met all of the required quality inspections.
|
|
Event Description
|
No additional information provided.
|
|