Model Number AVSM10040 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Restenosis (4576)
|
Event Date 07/18/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 12/2017).Device not returned.
|
|
Event Description
|
It was reported through the results of a clinical trial, that approximately seven months and two days post index procedure, the subject developed 70% of target lesion stenosis.Standard percutaneous transluminal angioplasty (pta) was used to successfully treat the target lesion.The current patient status was not provided.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned for evaluation and image was not provided for review.Therefore, the investigation is inconclusive for the reported issue.Pre dilation as well as post dilation was performed; there was no report of any irregularities or complication during initial stent placement.The definitive root cause could not be determined based upon available information.Labeling review: in reviewing the relevant labeling for this product, it was found that the instructions for use sufficiently address the potential risk.Based on the instructions for use complications and adverse events associated with the use of the covera vascular covered stent may include the usual complications associated with endovascular stent and covered stent placement and dialysis shunt revisions, which includes restenosis of the target vessel.Regarding pre and post dilation the instructions for use states: "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated" and "post dilate the covered stent with an angioplasty balloon sized appropriately as to ensure complete wall apposition to the reference vessel.Avoid balloon dilation in the healthy, non-stenosed segment of the vessel." h10: d4 (expiry date: 12/2017).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device not returned.
|
|
Event Description
|
It was reported through the results of a clinical trial, that approximately two years seven days post index procedure, the patient developed 30% of target lesion stenosis.A standard percutaneous transluminal angioplasty (pta) was used to successfully treat the target lesion.The current patient status was not provided.
|
|