Catalog Number 5C5479
Device Problem Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2021
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a patient was unable to disconnect the patient line of an amia automated pd cycler set from the transfer set.This was observed during use of the devices for peritoneal dialysis therapy.It was further reported that ¿the nurse cut the transfer set as it was not possible to disconnect¿.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
One actual sample was received for evaluation.The amia patient connector was attached to the female connector.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.The connection between the patient connector and female connector was performed with no issues noted.The patient connector was able to be hand removed from the female connector of the transfer set.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMIA AUTOMATED PD CYCLER SET
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key12484774
MDR Text Key271802133
Report Number1416980-2021-05782
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00085412153186
UDI-Public(01)00085412153186
Combination Product (y/n)Y
PMA/PMN Number
K151525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/17/2023
Device Catalogue Number5C5479
Device Lot NumberH21E17107
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2021
Date Manufacturer Received10/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MINICAP TRANSFER SET; MINICAP TRANSFER SET
Patient Age70 YR
-
-