MEDTRONIC NEUROMODULATION IMPLANTABLE NEUROSTIMULATOR; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number NEU_INS_STIMULATOR |
Device Problem
Loss of Data (2903)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Other relevant device(s) are: product id 8840 lot# serial# unknown , product type: programmer, physician.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare provider (hcp) who reported they encountered an issue with the clinician programmer erasing patient therapy settings during interrogation and getting the message ¿this application card is not compatible with the screening in session history stored in this device.¿ the hcp was redirected to contact the rep to get a tablet as it was reviewed with them that anytime they used a clinician programmer to interrogate a patient after they had been interrogated with a tablet, they would encounter this issue.
|
|