MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE; REVERSE 38MM HUMERAL LINER +0

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE; REVERSE 38MM HUMERAL LINER +0 Back to Search Results
Model Number 320-38-00
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Joint Dislocation (2374)
Event Date 08/27/2021
Event Type  Injury  
Manufacturer Narrative
300-30-08, equinoxe preserve stem 8mm.320-10-00, equinoxe reverse tray adapter plate tray +0.320-06-38, glenosphere 38mm.320-15-08, sup/post aug plate, r rs glenoid baseplate.
 
Event Description
Approximately 3 mos postop the initial r tsa, this (b)(6) y/o male patient felt a pop in his shoulder after moving it from the window of a vehicle.The shoulder was revised in outpatient surgery and was dismissed.The case report form indicates this event is definitely related to devices and procedure.This event report was received through clinical data collection activities.The devices will not return due to study policy.
 
Manufacturer Narrative
Section h10: (h3) based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the dislocation and subsequent revision cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition and the result of loosened supporting ligaments and muscles, which allowed for dislocation occurring due to the significant ligament reconstruction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Type of Device
REVERSE 38MM HUMERAL LINER +0
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key12488086
MDR Text Key271901100
Report Number1038671-2021-00486
Device Sequence Number1
Product Code KWT
UDI-Device Identifier10885862086655
UDI-Public10885862086655
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063569
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number320-38-00
Device Catalogue Number320-38-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age63 YR
Patient SexMale
Patient Weight68 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-