• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH CARDIOHELP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH CARDIOHELP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number CARDIOHELP-I
Device Problems Failure to Sense (1559); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/11/2021
Event Type  malfunction  
Manufacturer Narrative
Service of the affected cardiohelp unit was requested but is still pending.A follow-up emdr will be submitted when additional information becomes available.
 
Event Description
The customer reported that after several hours of patient support the flow reading values disappeared from the screen and the error message "art.Bubble sensor defective" was displayed.They replaced the flow/bubble sensor with another one and the error message "art.Bubble sensor disconnected" appeared.The cardiohelp was then replaced during treatment.No harm to patient reported.Complaint id: (b)(4).
 
Event Description
Complaint id: (b)(4).
 
Manufacturer Narrative
A flow/bubble sensor error message was reported.The failure occurred during treatment.A getinge service technician investigated the unit on 2021-10-19/21/22 and could confirm the failure.The sensor panel was replaced to restore function.The technician performed safety, calibration, and functionality checks to factory specifications.A sensor panel with the same failure was already investigated by getinge life-cycle-engineering and the supplier (b)(4).The most probable root cause was determined as the digiflow mini damaged by electro static discharge (esd).Based on the investigation results the reported failure "flow/ bubble sensor defective error" could be confirmed.The occurrence rate was calculated for the reported failure and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOHELP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key12488287
MDR Text Key271909405
Report Number8010762-2021-00506
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCARDIOHELP-I
Device Catalogue Number701072780
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/11/2021
Initial Date FDA Received09/17/2021
Supplement Dates Manufacturer Received02/09/2022
Supplement Dates FDA Received02/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-