WRIGHTS LANE SYNTHES USA PRODUCTS LLC UNK - VEPTR IMPLANTS: RIB HOOK; PROSTHESIS, RIB REPLACEMENT Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Skin Inflammation/ Irritation (4545)
Event Type  Injury  
Manufacturer Narrative
This report is for an unk - veptr implants: rib hook/unknown lot.Part and lot numbers are unknown; udi number is unknown.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This report is being filed after the review of a clinical evaluation report (cer) from a related research activity database (ddra) for 396 patients (209 female, 187 males) with early onset scoliosis primarily treated with veptr surgery beginning at <10 years of age.All complications data: 17 subjects experienced fatal complications, with zero related to veptr devices.-17 sae death, 1 sae -death -pneumonia secondary to covid-19, 3 sae- death ¿ respiratory complications 2 sae- death -chronic respiratory failure 2 sae- death- cardio- respiratory arrest 1 sae death-extended seizure 1 sae death- tracheal exsanguination 1 sae death- septic shock 6 sae death- unknown/died in sleep sae-surgical intervention -6 intraoperative (other complications) (n=174 study patients) 325 implant related complications , resolved by surgical intervention 182 implant related complications : surgical intervention, (n=325) ¿ planned surgery 143 implant related complications : surgical intervention, (n=325) ¿unplanned surgery -other complication: 63 pain 2 medical 287 wound related 12 neurologic injury/impairment 3 pseudarthrosis 84 pulmonary 183 other implant related complications (n=174 study patients) 329 implant related complications 293 implant related complications ,yes device related 36 implant related complications ,uncertain if device related 149/17 implant related complications (n=329) migration (no.Yes | no.Uncertain) 103/0 implant related complications (n=329) broken rods (no.Yes | no.Uncertain) 10/12 implant related complications (n=329) -dislodged hardware (no.Yes | no.Uncertain) 21/3 implant related complications (n=329) anchor prominence (no.Yes | no.Uncertain) 6/0 implant related complications (n=329) exposed hardware (no.Yes | no.Uncertain) 4/4 implant related complications (n=329)-skin erosion (no.Yes | no.Uncertain) this is for depuy synthes veptr and veptr ii.A copy of the clinical evaluation form is being submitted with this regulatory report.This report is for one (1) unk - veptr implants: rib hook.This is report 11 of 16 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK - VEPTR IMPLANTS: RIB HOOK
Type of Device
PROSTHESIS, RIB REPLACEMENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key12488978
MDR Text Key280601680
Report Number2939274-2021-05513
Device Sequence Number1
Product Code MDI
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial
Report Date 08/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/20/2021
Initial Date FDA Received09/17/2021
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-