Model Number 27074 |
Device Problems
False Alarm (1013); Battery Problem (2885); Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/19/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is currently being returned to resmed for an engineering investigation.The investigation methods, results, and conclusion are not finalized at this stage.When more information is available a supplemental report will be submitted.Resmed reference #: (b)(4).Device received, evaluation pending.
|
|
Event Description
|
It was reported to resmed that an astral device displayed a battery error message and an error message (sf131) related to the main blower temperature sensor.There was no patient harm or serious injury reported as a result of this incident.
|
|
Manufacturer Narrative
|
The astral device was returned to resmed for an investigation.Review of the device data logs confirmed the reported complaint.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the reported complaint was due to a software issue.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
|
|
Event Description
|
It was reported to resmed that an astral device displayed an error message (sf131) related to the main blower temperature sensor.There was no patient harm or serious injury reported as a result of this incident.
|
|