DAVIS & GECK CARIBE LTD ENDO CLIP III; CLIP, IMPLANTABLE Back to Search Results
Model Number 176630
Device Problems Mechanical Problem (1384); Device Slipped (1584); Mechanics Altered (2984)
Patient Problem Insufficient Information (4580)
Event Date 02/16/2019
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but three photos were available for evaluation.The first photo showed a clip applier and an unknown instrument, there were two unformed clips.The second photo showed a different angle of the two instruments, along with a fully formed clip and a malformed clip.The third photo showed some clips on tissue, the photo was not clear.It was reported that the clips did not load properly into the jaws as expected, did not close completely, malformed, and did not hold tightly to the vessel and detached.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during laparoscopic coelioscopy vesicle procedure, the surgeon used clip applier to clamp the cystic duct, however the handle got stuck and the clips did not hold and could be removed easily.Another device was used to complete the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP III
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key12501187
MDR Text Key272920304
Report Number9612501-2021-01554
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521057852
UDI-Public10884521057852
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K100242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model Number176630
Device Catalogue Number176630
Device Lot NumberJ8L1720Y
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/03/2021
Initial Date FDA Received09/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-