Model Number FT4 G3 |
Device Problem
High Test Results (2457)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Assays from different vendors can generate different values.This relates to the overall setups of the assays, the antibodies used, differences in reference materials/methods, and differences in the standardization methodology used.The investigation could not identify a product problem.
|
|
Event Description
|
The initial reporter stated they received questionable results for one patient sample tested with the elecsys ft4 iii assay on two cobas 8000 e 801 module analyzers.The results measured at the customer site were reported outside of the laboratory to a physician.The sample was initially tested on the customer's e 801 analyzer on (b)(6) 2021.The sample was repeated on an abbott architect analyzer.The sample was provided for investigation, where it was tested on a second e 801 analyzer on (b)(6) 2021.The sample was also treated with a 25% polyethylene glycol (peg) solution and tested.The serial number of the customer's e 801 analyzer was requested, but not provided.Ft4 reagent lot 52070100, with an expiration date of 31-jan-2022 was used on this analyzer.The serial number of the e 801 analyzer used for investigation is 14d9-06.Ft4 reagent lot 520701, with an expiration date of 31-jan-2022 was used on this analyzer.
|
|