Device evaluated by manufacturer: the device was returned for analysis.The pusher wire, introducer sheath and main coil were returned for this complaint.Visual inspection revealed the twist lock of the introducer sheath had been opened.The introducer sheath was inspected, and no anomalies were noted.The pusher wire was inspected and was found kinked.The main coil was found kinked and stretched.The proximal end has a smooth surface of the pusher wire and the interlocking arm was inspected, and no anomalies were noticed.For the main coil, the proximal end has a smooth surface.The interlocking arm was inspected, and it was detached (part was not returned).Dimensional inspection revealed that for the pusher wire, the overall dimension (od) of the distal solder joint, mid solder joint, distal tfe and proximal tfe matches with the specification.For the main coil, the zap tip (od) and primary coil (od) are within the specification, however, the number of fiber bundles were lacking.
|
Reportable based on device analysis completed on 30aug2021.It was reported that device was stuck in the catheter.The target lesion was located in the splenic artery.A 3mm x 6cm interlock was selected for use.During the procedure, the coil got stuck in the catheter and prematurely deployed in the protective sheath.The device simply pulled out and the procedure was completed with another of the same.A total of 2 coils were used in the procedure.There were no complications reported and the patient is stable.However, device investigations revealed that the interlocking arm was detached and there were missing fiber bundles.
|