Correction / additional information: d9: date returned to manufacturer.E1: region, telephone and fax.G2: distributor.G4: 510k number.H6: codes.Investigation results: visual investigation: the investigation was carried out visually and microscopically with the digital-camera.Here broken off noses, broken off latches and clip jams were found.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 3(5) x probability of occurrence 3(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a psc (b)(4) has been initiated.Any action regarding capa will be addressed with this case.
|