• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TOGGLELOC WITH ZIPLOOP INLINE; PROSTHESIS SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TOGGLELOC WITH ZIPLOOP INLINE; PROSTHESIS SHOULDER Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign: india.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that while pulling the button from femoral side the sutures snapped out.Attempts have been made and there is no further information at this time.
 
Manufacturer Narrative
Visual evaluation of the returned device showed it had broken sutures and a damaged button.Sharp edges were identified, however, it is unknown if this damage was sustained during implanting or explanting.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOGGLELOC WITH ZIPLOOP INLINE
Type of Device
PROSTHESIS SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12512599
MDR Text Key272689192
Report Number0001825034-2021-02680
Device Sequence Number1
Product Code JDR
UDI-Public(01)0088030456763417)25029(10)512950
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K130033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110005087
Device Lot Number512950
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/28/2021
Initial Date FDA Received09/22/2021
Supplement Dates Manufacturer Received10/25/2021
Supplement Dates FDA Received11/10/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/29/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age36 YR
Patient SexMale
-
-