EDWARDS LIFESCIENCES PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV/VIP THERMODILUTION CATHETER; CATHETER, OXIMETER, FIBER-OPTIC Back to Search Results
Model Number 777F8
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2021
Event Type  malfunction  
Manufacturer Narrative
One 777f8 catheter without any attached components was returned for examination.The reported event of balloon issue was confirmed.The balloon was found to be torn around circumference at both the proximal and distal bonding sites.The torn balloon piece was not returned.All through lumens were patent without any leakage or occlusion.No visible damage or inconsistency was observed from the catheter body.Visual examination was performed under microscope at 20x magnification and unaided eye.A review of the manufacturing records indicated that the product met specifications upon release.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the balloon did not inflate when the customer inflated the balloon on the first day of use.It is unknown if the balloon test before use was performed, or any balloon issue was detected before use there were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV/VIP THERMODILUTION CATHETER
Type of Device
CATHETER, OXIMETER, FIBER-OPTIC
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key12513512
MDR Text Key273112368
Report Number2015691-2021-05295
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/08/2022
Device Model Number777F8
Device Catalogue Number777F8
Device Lot Number63411142
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2021
Was the Report Sent to FDA? No
Date Manufacturer Received09/03/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/11/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-