Catalog Number 2C4711K |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/27/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the ¿rubber membrane¿ (bladder) of a folfusor sv (small volume) 2.5 ml was torn at the time of filling.The device was filled with fluorine and glucose.This was identified prior to use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The actual device was not available; however, a photograph of the sample was provided for evaluation.The returned photograph was reviewed, and it was noted that the photograph showed evidence of a bladder rupture.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|