One 4 lesion nt2000" pain management rf generator was received for analysis.Visual inspection of the returned rf generator indicated all labels are intact, legible and are free of physical damage.It was also verified that all internal components were secured, and normal wear was noted on the chassis exterior.Ac power was applied to the rf generator and a successful power on self-test was executed.Inspection of the internal components revealed an electrically failed and discolored capacitor at location c49 on the radio frequency control board.Based on the information provided to abbott and the investigation performed, the root cause of the field reported issue was isolated to abnormal functionality of the radio frequency control board.
|