• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. 4 LESION NT2000 PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. 4 LESION NT2000 PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY Back to Search Results
Model Number RFG-NT-2000
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/30/2021
Event Type  malfunction  
Manufacturer Narrative
One 4 lesion nt2000" pain management rf generator was received for analysis.Visual inspection of the returned rf generator indicated all labels are intact, legible and are free of physical damage.It was also verified that all internal components were secured, and normal wear was noted on the chassis exterior.Ac power was applied to the rf generator and a successful power on self-test was executed.Inspection of the internal components revealed an electrically failed and discolored capacitor at location c49 on the radio frequency control board.Based on the information provided to abbott and the investigation performed, the root cause of the field reported issue was isolated to abnormal functionality of the radio frequency control board.
 
Event Description
This event is being reported due to an electrically failed and discolored capacitor of the generator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4 LESION NT2000 PAIN MANAGEMENT RF GENERATOR
Type of Device
GENERATOR, LESION, RADIOFREQUENCY
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12520470
MDR Text Key272950837
Report Number2184149-2021-00296
Device Sequence Number1
Product Code GXD
UDI-Device Identifier05415067022417
UDI-Public05415067022417
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111576
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRFG-NT-2000
Device Catalogue NumberRFG-NT-2000
Device Lot Number7496791
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/07/2020
Is the Device Single Use? No
Type of Device Usage N
Removal/Correction Number2184149-04/22/21-001-C
Patient Sequence Number1
-
-