|
Model Number REPLY DR |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 03/22/2021 |
Event Type
Injury
|
Event Description
|
Reportedly, the subject pacemaker was implanted on (b)(6) 2011 to treat an avb.The patient was hospitalized from (b)(6) to (b)(6) 2021 to replace the pacemaker and leads, which were not operating correctly.During the previous follow-ups, a dysfunction of the atrial lead was observed (increase of the lead impedance and loss of capture), then a similar issue was observed on the ventricular lead.The ecg performed before the re-intervention was normal (sinus rhythm, with normal pr interval, avb).On (b)(6) 2021, upon explantation of the pacemaker, a complete fracture of the header was observed, explaining the dysfunctioning of the leads.The leads were tested with a psa.Proper operation of the leads was observed with thresholds below 1v.As a result, the leads were kept implanted and connected to a new pacemaker.
|
|
Manufacturer Narrative
|
Preliminary of analysis of the returned device showed that its header was detached from the can.
|
|
Event Description
|
Reportedly, the subject pacemaker was implanted on (b)(6) 2011 to treat an avb.The patient was hospitalized from (b)(6) 2021 to replace the pacemaker and leads, which were not operating correctly.During the previous follow-ups, a dysfunction of the atrial lead was observed (increase of the lead impedance and loss of capture), then a similar issue was observed on the ventricular lead.The ecg performed before the re-intervention was normal (sinus rhythm, with normal pr interval, avb).On (b)(6) 2021, upon explantation of the pacemaker, a complete fracture of the header was observed, explaining the dysfunctioning of the leads.The leads were tested with a psa.Proper operation of the leads was observed with thresholds below 1v.As a result, the leads were kept implanted and connected to a new pacemaker.It should be noted that the patient suffered a fracture in 2015.Preliminary analysis of the available patient files showed unstable and high ventricular lead impedances (in the order of 3000 ohms) since (at least) (b)(6) 2020.Loss of ventricular capture was observed in the real time test from (b)(6) 2021.Episodes with non-physiological signals were noted in both channels since (at least) (b)(6) 2021.An intermittent behavior was identified relative to the ventricular channel.
|
|
Manufacturer Narrative
|
B5 updated.H6 (type of investigation) updated.Review of the device history records showed no anomaly.
|
|
Event Description
|
Reportedly, the subject pacemaker was implanted on (b)(6) 2011 to treat an avb.The patient was hospitalized from (b)(6) 2021 to replace the pacemaker and leads, which were not operating correctly.During the previous follow-ups, a dysfunction of the atrial lead was observed (increase of the lead impedance and loss of capture), then a similar issue was observed on the ventricular lead.The ecg performed before the re-intervention was normal (sinus rhythm, with normal pr interval, avb).On (b)(6) 2021, upon explantation of the pacemaker, a complete fracture of the header was observed, explaining the dysfunctioning of the leads.The leads were tested with a psa.Proper operation of the leads was observed with thresholds below 1v.As a result, the leads were kept implanted and connected to a new pacemaker.It should be noted that the patient suffered a fracture in 2015.Preliminary analysis of the available patient files showed unstable and high ventricular lead impedances (in the order of 3000 ohms) since (at least) (b)(6) 2020.Loss of ventricular capture was observed in the real time test from (b)(6) 2021.Episodes with non-physiological signals were noted in both channels since (at least) (b)(6) 2021.An intermittent behavior was identified relative to the ventricular channel.
|
|
Event Description
|
Reportedly, the subject pacemaker was implanted on (b)(6) 2011 to treat an avb.The patient was hospitalized from (b)(6) 2021 to replace the pacemaker and leads, which were not operating correctly.During the previous follow-ups, a dysfunction of the atrial lead was observed (increase of the lead impedance and loss of capture), then a similar issue was observed on the ventricular lead.The ecg performed before the re-intervention was normal (sinus rhythm, with normal pr interval, avb).On (b)(6) 2021, upon explantation of the pacemaker, a complete fracture of the header was observed, explaining the dysfunctioning of the leads.The leads were tested with a psa.Proper operation of the leads was observed with thresholds below 1v.As a result, the leads were kept implanted and connected to a new pacemaker.It should be noted that the patient suffered a fracture in the tibia in 2015.Preliminary analysis of the available patient files showed unstable and high ventricular lead impedances (in the order of 3000 ohms) since (at least) (b)(6) 2020.Loss of ventricular capture was observed in the real time test from (b)(6) 2021.Episodes with non-physiological signals were noted in both channels since (at least) (b)(6) 2021.An intermittent behavior was identified relative to the ventricular channel.
|
|
Manufacturer Narrative
|
B5 updated.Please refer to the attached analysis report.
|
|
Search Alerts/Recalls
|
|
|