Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/30/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
|
Event Description
|
It was reported that the solasic catheters were too thin and stretched out, so they were loose in the statlock.The user had to stuff toilet paper or tissue in it.Per follow up information received on 05sep2021, stated the foley catheter was too thin and appeared to be stretched with white stretch markings just below where the sample ports connection junction was located.When the user placed the catheter into the statlock, it slides and pulls through so they used toilet paper or tissue to prevent this from happening.
|
|
Manufacturer Narrative
|
The reported event was inconclusive because no sample was returned for evaluation.A potential root cause for this failure could be due to "incorrect catheter buildup and/or finish dipping".It was unknown whether the device had met specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.A dhr review could not be performed without a lot number.The product catalog number and the lot number for this device are unknown.Therefore, bd is unable to determine the associated labeling to review.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
|
|
Event Description
|
It was reported that the solasic catheters were too thin and stretched out, so they were loose in the statlock.The user had to stuff toilet paper or tissue in it.Per follow up information received on 05sep2021, stated the foley catheter was too thin and appeared to be stretched with white stretch markings just below where the sample ports connection junction was located.When the user placed the catheter into the statlock, it slides and pulls through so they used toilet paper or tissue to prevent this from happening.
|
|
Search Alerts/Recalls
|