A facility reported that during a posterior cervical fusion procedure, the patient's head slipped while supported by the mayfield skull clamp (a3059) and sustained scalp lacerations.The surgeon switched to a different equipment to complete the procedure.Additional information received indicates that 80 pounds of pressure was applied during the procedure.Additionally, the patient was positioned prone and after this issue, the patient was repositioned and switched to gardner wells tongs to support the head and surgery was completed with no additional issues reported.Surgical delay of 30 minutes was reported.
|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10.The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: with respect to the returned unit, the unit passed all specific functional testing requirements and will be returned to the customer without requiring repair.When the unit is properly positioned and put under pressure, the unit would not have slipped.Root cause analysis: evaluation found no device deficiencies that would have contributed to the reported complaint.Repairs could not duplicate slippage.When unit is properly positioned and put under pressure, the unit would not have slipped.Probable root cause for reported incident is improper placement of the skull clamp.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|