• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL NEONATAL FEEDING TUBE WITH ENFIT CONNECTOR; DH CPK NG TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL NEONATAL FEEDING TUBE WITH ENFIT CONNECTOR; DH CPK NG TUBES Back to Search Results
Model Number 41-1155
Device Problem Material Split, Cut or Torn (4008)
Patient Problem Discomfort (2330)
Event Date 08/30/2021
Event Type  malfunction  
Manufacturer Narrative
A review of the device history record is not possible as no lot number was provided.The actual complaint product was returned for evaluation.The root cause of the reported issue could not be conclusively determined.However; this seems to be a user related problem since as per the ifu, vigorous syringe force should not be used to irrigate, administer liquids, or unblock the tube.All information reasonably known as of 24 sep 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).
 
Event Description
It was reported that the ng [nasogastric] tube was "broken." according to the report, the "patient [was] fussy all day with feeds, kub [kidney-ureter-bladder x-ray] obtained which showed that ngt was broken in patient's stomach (distal tip hanging on by a thread).Ngt pulled and replaced." no injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NASOGASTRIC/NASOINTESTINAL NEONATAL FEEDING TUBE WITH ENFIT CONNECTOR
Type of Device
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp 84048
MX   84048
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key12530352
MDR Text Key273341088
Report Number9611594-2021-00137
Device Sequence Number1
Product Code KNT
UDI-Device Identifier00350770460604
UDI-Public00350770460604
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K831328
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number41-1155
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/31/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-