HAEMONETICS CORPORATION CELL SAVER ELITE PLUS AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE SET - 225ML Back to Search Results
Model Number CSE-P-225-JA
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2021
Event Type  malfunction  
Manufacturer Narrative
On august 27, 2021 haemonetics manufacturing performed a visual inspection of the received bowl from the cell saver® elite set - 225ml and confirmed blood in the inner core as well as 6 cracks in the inner core base.Although there was no serious injury or harm, past reporting indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due the event of the past report.
 
Event Description
On (b)(6) 2021 haemonetics was notified of a inner core leak which was observed during a procedure in (b)(6), utilizing the cell saver® elite® autotransfusion system and cell saver® elite set - 225ml.There was no reported impact to patients' health.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER ELITE PLUS AUTOTRANSFUSION SYSTEM
Type of Device
CELL SAVER ELITE SET - 225ML
Manufacturer (Section D)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer Contact
janice kiser
125 summer street
boston, MA 02110
MDR Report Key12530679
MDR Text Key279936461
Report Number1219343-2021-00126
Device Sequence Number1
Product Code CAC
UDI-Device Identifier10812747016544
UDI-Public(01)10812747016544(17)231005(10)1020041
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120586
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/05/2023
Device Model NumberCSE-P-225-JA
Device Lot Number1020041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/27/2021
Initial Date FDA Received09/24/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/05/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-