• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOCLEAN, INC. SOCLEAN 2; SLEEP EQUIPMENT MAINTENANCE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOCLEAN, INC. SOCLEAN 2; SLEEP EQUIPMENT MAINTENANCE SYSTEM Back to Search Results
Model Number SC1200
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Dyspnea (1816); Skin Inflammation/ Irritation (4545); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 05/13/2020
Event Type  Injury  
Manufacturer Narrative
File a 30-day mdr.An assessment was conducted.The event is still considered reportable under fda's regulation.Soclean has reviewed and made improvements to our complaint handling procedure to ensure compliance with fda regulations.These improvements were recommended as a result of an internal audit conducted by an external independent regulatory consultant.We revised our decision trees and re-evaluated retrospective complaints to ensure reportable events were correctly identified.Per fdas guidance soclean is now reporting these retrospective mdrs.In addition, soclean has opened several capa's to remediate the audit findings.
 
Event Description
Customer reports nose irritation and breathing difficulty with md intervention requiring antibiotics and claritin.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOCLEAN 2
Type of Device
SLEEP EQUIPMENT MAINTENANCE SYSTEM
Manufacturer (Section D)
SOCLEAN, INC.
12 vose farm road
peterborough NH 03458
Manufacturer (Section G)
SOCLEAN, INC.
12 vose farm road
peterborough NH 03458
Manufacturer Contact
anne nadeau
12 vose farm road
peterborough, NH 03458
6033712533
MDR Report Key12530837
MDR Text Key273302552
Report Number3009534409-2021-00188
Device Sequence Number1
Product Code LRJ
UDI-Device Identifier00187293000860
UDI-Public187293000860
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 07/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberSC1200
Device Catalogue NumberSC1200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2020
Initial Date Manufacturer Received 05/13/2020
Initial Date FDA Received09/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-