It was reported that the patient had a reaction to the comfort hard-soft splint.It was reported that the patient had swollen lymph nodes, extreme fatigue, and brain fog.The patient initial use of the device was (b)(6) 2021 and noticed the reaction (b)(6) 2021 and progressively got worse.The patient discontinued usage of the device on (b)(6) 2021.Once discontinuation, the reaction lasted about a week.The patient did not receive any medical treatment.There is no medical history noted and has no known allergies.The patient current status is not as "feeling back to normal." with regard to the device: the provider noted that the device was cleaned but was not specific.The patient rinsed with mild soap, water and cleaned with toothbrush.
|
The device has not been returned.However, the non-visual device investigation has been completed and the results are as follows: dhr results: the dhr was reviewed and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.Supplier (erkodent) reviewed the associated material lot and confirmed no manufacturing deviation or abnormality.Lot# e-pro 4.0-11615 (erkoloc-pro) was manufactured from april 16, 2021 and was assigned an expiration of april 2024 stock product reviewed results: no stock product was available for review since the device was fabricated per physician's prescription only.Investigation methods/results: no device has been returned from the customer.There have been attempts to establish communication with the provider but there has been no response.However, the non-visual device investigation has been completed.Root cause: a root cause for this complaint cannot be explicitly determined.Ifu 9091 rev 4.0 (comfort h/s bite splint instruction for use) states "brush and floss your teeth before use.Rinse mouth well with clean water before inserting the device.If patient uses mouthwash, all traces of mouthwash should be removed by thoroughly rinsing out mouth with water.Rinse bite splint well with clean, cool water before and after use.Clean bite splint with clean, cool water only and let air dry." ifu provides warning "do not clean or soak in mouthwash; do not use denture cleanser, hot water, alcohol, hydrogen peroxide; do not place in direct sunlight".It is possible that reactions could be caused by mouthwash, toothpaste, or soaking material.Per the reported information, the customer rinsed with mild soap, water and cleaned with toothbrush.Supplier erkodent reviewed the incident details and determined an allergic reaction cannot be ruled out.Glidewell research team and namsa conducted a series of testing on a similar thermoformed sleep device following iso 10993 (biological evaluation of medical devices) and the device was evaluated for potential cytotoxicity, skin irritation, delayed dermal contact sensitization and oral mucosal irritation.The test article was thermoformed with layers of erkodent material (erkoloc-pro and erkodur).The test results were listed below and summarized in biocompatibility report for sleep device (rpt 9733 rev 1.0) · for cytotoxicity testing, the test article extract showed no evidence of causing cell lysis or toxicity.· for skin irritation, there was no erythema and no edema observed on the skin of the animals treated with the test article.· for sensitization testing, the test article extracts showed no evidence of causing delayed dermal contact sensitization.· the test article showed nonirritant to the oral mucosa as compared to the control article.The device materials have been found to be biocompatible through the testing.There was no cytotoxic, sensitization, skin irritation, or oral mucosal irritation found in any of the test articles.
|