It was reported that the patient presented in clinic for generator implant procedure.During the procedure, it was found that upon connecting the atrial lead to the new implantable cardioverter defibrillator (icd), the obtained lead measurements were incorrect and did not match the results obtained from testing using the pacing system analyzer (psa).It was then attempted to disconnect the lead from the header.It was reported that the set screw had become stripped due to excessive force from the torque wrench.Eventually, the lead was able to be removed with excess force.The procedure was completed using an alternate icd on (b)(6) 2021.There were no patient consequences.
|