• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FMS VUE PUMP; DISTENSION UNIT, FLUID, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FMS VUE PUMP; DISTENSION UNIT, FLUID, ARTHROSCOPIC Back to Search Results
Model Number 284002
Device Problem Overfill (2404)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2021
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported by the sales rep that the fill chamber light on the fms vue pump device did not light up once tubing was connected and then it overflowed and began running erratically.There was no procedure nor patient involvement reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device was received at the service center and evaluated.It was reported that ¿sales consultants were testing out facility's malfunctioned pump and came across another pump having issues.Fill chamber light did not light up once tubing was connected.Fill chamber overflowed and began running erratically.¿ per service reports, no defect was found with the device, therefore this complaint cannot be confirmed.As the reported problem was not confirmed and no defects identified, a root cause for the issue that was experienced by the user cannot be determined.No manufacturing record evaluation is required as no manufacturer or service related issues were found with the device.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
MDR Report Key12536299
MDR Text Key273572458
Report Number1221934-2021-02810
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020591
UDI-Public10886705020591
Combination Product (y/n)N
PMA/PMN Number
K130169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/23/2017
Device Model Number284002
Device Catalogue Number284002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2021
Initial Date Manufacturer Received 09/14/2021
Initial Date FDA Received09/27/2021
Supplement Dates Manufacturer Received11/04/2021
Supplement Dates FDA Received11/04/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-