ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Difficult to Remove (1528); Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2021
Event Type  malfunction  
Event Description
It was reported a patient presented for a procedure on (b)(6) 2021.During left ventricular lead implant, the guidewire became stuck within the lead so the lead was not used.Post-procedure the patient was stable.
 
Manufacturer Narrative
Correction: h6: should have been difficult to remove rather than failure to disconnect.
 
Manufacturer Narrative
The reported event of unable to remove guidewire was confirmed.As received, a complete lead was returned in one piece.Visual inspection found the guidewire stuck inside the lead.Unable to perform the guidewire removal test due to the overtorqued inner coil inside the connector region as received.The cause of the reported event was isolated to the bunching up of inner coil inside the connector region and stripped ptfe coating of guidewire.
 
Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12538092
MDR Text Key273532261
Report Number2017865-2021-31984
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberS000079495
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-