The quality check performed after the bed was returned to the service center confirmed that the bed was in full working condition.No malfunction was identified.Based on the collected information the root cause of the investigated event cannot be established.No details regarding circumstances of the fall were possible to be obtained from the customer.The instructions for use for citadel patient care system (830.238-en rev.G) takes into consideration the risk of patient's fall/inadvertent bed exit on several occasions: "specialty surfaces have different shear and support characteristics than conventional surfaces and may increase the risk of patient (.) inadvertent bed exit.Monitor patients frequently to guard against patient entrapment." "to minimize the risk of falls or injury, the bed should always be in the lowest practical position when the patient is unattended." "whether and how to use side rails or restraints is a decision that should be based on each patient's needs and should be made by the patient and the patient's family, physician and caregivers, with facility protocols in mind.Caregivers should assess risks and benefits of side rail/restraint use (including entrapment and patient falls from bed) in conjunction with individual patient needs, and should discuss use or non-use with patient and/or family.Consider not only the clinical and other needs of the patient but also the risks of fatal or serious injury from falling out of bed (.).It is recommended that side rails (if used) be locked in the full upright position when the patient is unattended." arjo device was used for a patient treatment when the event occurred and from that perspective it played a role in the event.The device was performing as intended.No malfunction was found during the device evaluation.This complaint is deemed reportable due to allegation of patient's fall from the bed.
|