• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HAEMONETICS CORPORATION CELL SAVER ELITE AUTOTRANSFUSION SYSTEM; CS ELITE PROCESSING KIT, 225ML

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAEMONETICS CORPORATION CELL SAVER ELITE AUTOTRANSFUSION SYSTEM; CS ELITE PROCESSING KIT, 225ML Back to Search Results
Model Number CSE-P-225
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/03/2021
Event Type  malfunction  
Manufacturer Narrative
Due to patient confidentiality, the customer was not able to provide patient information or give any details of patients status.Haemonetics field service engineer inspected the device and there were no faults found.On june 11, 2021 haemonetics manufacturing performed a visual inspection of the received bowl from the cell saver® elite set - 125ml and confirmed blood in the inner core with a crack in the inner core base.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due the event of the past report.
 
Event Description
On (b)(6) 2021, haemonetics was notified of air detected early error codes (122, 130, 280, 161, 115) and a required 9 unit transfusion which occurred during a procedure in the (b)(6), utilizing the cell saver® 5 autologous recovery system and cell saver® 5 set - 225ml.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER ELITE AUTOTRANSFUSION SYSTEM
Type of Device
CS ELITE PROCESSING KIT, 225ML
Manufacturer (Section D)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer Contact
janice kiser
125 summer street
boston, MA 02110
MDR Report Key12540554
MDR Text Key273580836
Report Number1219343-2021-00122
Device Sequence Number1
Product Code CAC
UDI-Device Identifier10812747016537
UDI-Public(01)10812747016537(17)231123(10)0121024
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K120586
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 06/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/23/2023
Device Model NumberCSE-P-225
Device Lot Number1120074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2021
Initial Date Manufacturer Received 06/11/2021
Initial Date FDA Received09/28/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/23/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-